Electronic instructions for use (eIFU) are legal today for the large majority of professional-use medical devices in the European Union, the United Kingdom, Switzerland, the United States, Canada, Australia, Brazil, Saudi Arabia and Japan. The rules differ in the details, but they share one shape: eIFU is permitted for devices used by healthcare professionals, paper remains mandatory for devices used by lay persons, and every manufacturer must be able to send a printed copy on request. This article reviews the legal status of an eIFU implementation market by market, with the primary references at the end, as of November 2026.

In regulatory texts, electronic instructions for use means instructions provided instead of paper: on a website, on electronic media supplied with the device, or on a display built into the device. Supplementary digital content (a QR code that leads to a video, an app, a PDF that duplicates the paper IFU) is not an eIFU in the legal sense and does not need a specific permission. What needs permission is the decision to stop putting paper in the box. That is the decision this article is about.
Legal basis. Regulation (EU) 2017/745 (MDR), Annex I, section 23.1(f), allows instructions for use to be provided in non-paper form under the conditions set by an implementing act. That act is Commission Implementing Regulation (EU) 2021/2226, in application since 4 January 2022 and amended by Commission Implementing Regulation (EU) 2025/1234, in force since 16 July 2025 [1][2].
Which devices. Until July 2025, eIFU was limited to implantable devices, fixed installed devices, devices with a built-in display and their accessories. Regulation 2025/1234 removed that list. Article 3(1) of the amended regulation now allows eIFU for all devices covered by Article 1(4) MDR that are intended for professional users: every class of medical device, their accessories and the Annex XVI products without an intended medical purpose. Recital 3 confirms that legacy devices under the Article 120 transitional provisions are included [2][3].
The lay-user rule. Article 3(2) is the one line that still forces paper: where it is reasonably foreseeable that a professional-use device will also be used by lay persons, the instructions intended for lay persons must be provided in paper form. Home-use devices stay on paper.
Conditions that still apply. The amendment did not relax the operating conditions of 2021/2226, which are the basis of any eIFU platform:
Regulation (EU) 2017/746 (IVDR) does not have an implementing regulation on eIFU and Regulation 2021/2226 does not apply to IVDs. The rule sits directly in Annex I, section 20.1(f): instructions for use may be provided in non-paper form for devices intended for professional use only, except devices intended for near-patient testing. Self-testing devices, which are by definition not professional-use, remain on paper [5]. We argued in January that the near-patient exclusion no longer matches clinical practice (see our earlier post); as of November 2026 the text is unchanged.
Switzerland applies the EU implementing regulations directly. Under Article 95 of the Medical Devices Ordinance (MedDO, SR 812.213), Swissmedic adopted Regulation 2021/2226 as amended by 2025/1234 with immediate effect on 8 August 2025, so the same scope (all professional-use devices, including legacy devices) and the same conditions apply in Switzerland, including the obligation to record the eIFU address in the UDI database [6].
Great Britain. The UK Medical Devices Regulations 2002 (UK MDR 2002), regulation 4J, carry forward the pre-2021 EU scope: eIFU is permitted for active implantable and implantable devices, fixed installed devices, devices with a built-in display of the instructions, and standalone software, for professional users only and subject to a risk assessment and paper on request. The 2025 EU extension to all professional-use devices has not been mirrored in Great Britain. IVDs follow MEDDEV 2.14/3 rev.1. The MHRA guidance on eIFU remains the reference; the wider reform of the pre-market regulations is in draft and should be watched for a change of scope [7].
Northern Ireland. EU MDR and IVDR continue to apply under the Windsor Framework, so Regulation 2021/2226 as amended by 2025/1234 applies to devices placed on the Northern Ireland market.
The Federal Food, Drug, and Cosmetic Act, section 502(f), as amended by section 206 of the Medical Device User Fee and Modernization Act of 2002 (MDUFMA), allows the labelling of prescription devices intended for use in health care facilities to be provided solely by electronic means, provided that the labelling complies with all applicable requirements and that the user can request paper labelling, which must be provided promptly and at no additional cost. The Medical Devices Technical Corrections Act of 2004 (MDTCA) widened the provision: electronic labelling is now also allowed for prescription devices used by a health care professional regardless of the setting, for in vitro diagnostic devices used by health care professionals and for IVDs used in blood establishments. Over-the-counter devices and prescription devices used at home by lay persons remain on paper [16]. FDA's implementing memo adds that the conversion from paper to electronic does not by itself require a PMA supplement or a new 510(k), that verification and validation of the electronic labelling belong in the design history file, and that submissions must identify the website and confirm that its content is identical to the paper version. The content requirements of 21 CFR Part 801 (devices) and Part 809 (IVDs) continue to apply [8][9].
The Medical Devices Regulations (SOR/98-282), section 21(2), allow the information required by section 21(1), including the directions for use, to be provided electronically for devices that are not sold to the general public. Health Canada's labelling guidance specifies that the electronic label or the web address must accompany the device at the time of sale or delivery, that the electronic content must be identical to the paper version (an attestation letter from a senior official is suggested), and that a paper copy must be supplied as soon as possible on request. Devices sold to the general public fall under sections 22 and 23: directions for use on or in the package, in both official languages [10].
The Therapeutic Goods Administration permits eIFU for medical devices that are not supplied to the general public, under the Essential Principles of the Therapeutic Goods (Medical Devices) Regulations 2002 and its guidance Supplying electronic instructions for use (eIFU) for medical devices. The manufacturer must document a risk assessment, provide the instructions on a website or electronic media, and supply a paper copy on request without delay or cost. Devices supplied to the general public and IVD medical devices must include paper instructions; electronic instructions can be offered in addition. The TGA consulted in 2024 on allowing more flexible IFU formats and the government has approved the preparation of regulatory amendments, so the scope may widen; as of November 2026 the 2018 guidance still applies [11].
ANVISA's RDC 751/2022, which replaced RDC 185/2001 and consolidated labelling and IFU rules, contains a dedicated section on instructions for use in non-printed format. The manufacturer or registration holder is responsible for making the electronic IFU available (a reference to ANVISA's own portal is not enough); the current version must be uploaded to ANVISA's medical device document repository (Article 33); and the outer label must indicate how to obtain the corresponding IFU, for example through a download address or a QR code (Article 51). As under the previous rules, devices for home use by lay persons and IVDs for self-testing remain on paper [12].
The Saudi Food and Drug Authority's Requirements for Electronic Instructions for Use (eIFU) of Medical Devices (2020) permit eIFU for medical devices and IVDs intended for use by healthcare professionals. Devices for self-use and point-of-care IVDs are excluded. The manufacturer must complete and maintain a risk assessment, state on the labelling that the IFU is electronic, display release date, jurisdiction and version in the document, retain versions for at least 2 years after the expiry of the last device (15 years for devices without an expiry date), and host the files on a website without login, with adequate cybersecurity and minimal downtime [13].
Japan went further than any other market. Following the 2019 amendment of the Pharmaceuticals and Medical Devices Act, electronic package inserts became the default for prescription medicines, medical devices and IVDs from 1 August 2021 (with a two-year transition). Users scan the GS1 barcode on the outer box with the PI-navi app to open the current package insert; marketing authorisation holders must publish the insert on PMDA's website and register it in the Safety Information Posting System. Products bought directly by consumers keep paper inserts [14].
Under Article 33 of the Medical Devices Act, the Taiwan FDA has announced the device items for which an electronic manual may replace the Chinese-language paper manual. Permitted formats are built-in help systems, portable electronic media and the manufacturer's website; the label or package must explain how to obtain a paper manual and give the manufacturer's contact details. Check the current announcement for the item list before relying on it [15].
| Market | Legal basis | Professional-use devices | Lay / home-use devices | Distinctive condition |
|---|---|---|---|---|
| EU (MDR) | Reg. 2021/2226 as amended by 2025/1234 | All devices incl. Annex XVI and legacy | Paper | eIFU URL in Eudamed UDI database; paper within 7 days; 10/15-year retention |
| EU (IVDR) | IVDR Annex I 20.1(f) | Yes, except near-patient testing | Paper (self-tests) | No implementing act; conditions by analogy with 2021/2226 |
| Switzerland | MedDO Art. 95; Swissmedic 8 Aug 2025 | As EU | Paper | Same as EU |
| Great Britain | UK MDR 2002 reg. 4J; MHRA guidance | Implantables, fixed installed, built-in display, software only | Paper | Pre-2021 scope; reform pending |
| Northern Ireland | EU MDR/IVDR | As EU | Paper | As EU |
| United States | FD&C Act 502(f) (MDUFMA s.206, MDTCA 2004) | Rx devices used by health care professionals or in health care facilities; professional-use IVDs | Paper (OTC, home-use Rx) | Paper promptly on request, no cost; V&V in DHF |
| Canada | MDR SOR/98-282 s.21(2) | Devices not sold to the general public | Paper, both official languages | Identical content; attestation; paper ASAP |
| Australia | Essential Principles; TGA eIFU guidance | Devices not supplied to the general public (not IVDs) | Paper (electronic in addition); IVDs on paper | Risk assessment; paper without delay or cost; reform in preparation |
| Brazil | RDC 751/2022 (Arts. 33, 51) | Yes | Paper (home use, self-test IVD) | Upload to ANVISA repository; label states how to obtain IFU |
| Saudi Arabia | SFDA eIFU Requirements (2020) | Devices and IVDs for HCP use | Paper (self-use, POC IVD) | Version, date and jurisdiction on the document; no login |
| Japan | PMD Act (amended 2019), from 1 Aug 2021 | Electronic by default (Rx drugs, devices, IVDs) | Paper for consumer products | GS1 barcode, PI-navi, PMDA posting |
| Taiwan | Medical Devices Act Art. 33 announcement | Listed items only | Paper | Label must state how to get paper |
Whatever the market, an eIFU platform has to cover the same seven obligations: a documented risk assessment; an always-available, secure website with no login; every version kept with its publication date for 10 to 15 years; a paper-copy process with a measurable response time; the eIFU indicator and access information on the label; a stable URL that can be registered in UDI databases; and a validation file that shows the system does what the regulations say. A home-built intranet page or a PDF on the corporate site rarely meets all seven, which is why most manufacturers now use a dedicated service.
eLeaflet™ by myHealthbox is built for exactly these obligations: a validated platform (EU GMP Annex 11, FDA 21 CFR Part 11) that publishes IFUs in 45 languages on the global eIFU search engine and, if you prefer, on your own company address, with version history, e-signature approval, QR and GS1 Digital Link access on the pack, and a built-in paper-copy request. If your IFUs are still in the box, request a demo and we will show you what going electronic involves for your product range.
This article is a regulatory overview for information purposes, current as of 9 November 2026. It is not legal advice. Check the official texts, which are linked above, before taking a product decision.